bbraun vulnerabilities, newest first
Filter
By year: all 2026 2025 2024 2023 2022 2021
Includes CVEs the NVD has not named yet, filed by their publishers under: b. braun melsungen ag.
- 8.5CVE-2026-97361EPSS 0.1%
There exists an improper privilege management vulnerability that could cause privilege escalation at the SpaceCom and Space Battery pack SP with WiFi.
- 7.0CVE-2026-97358EPSS 0.5%
An inclusion of web functionality from an untrusted source could allow remote code execution on SpaceCom
- 6.9CVE-2026-97357EPSS 0.3%
An Out-Of-Bounds Write due to missing input validation is possible in the autoprogramming interface of SpaceCom and Space Battery pack with WiFi which could lead to a denial of service of the affected communication device
- 9.8CVE-2025-3365EPSS 0.6%
A missing protection against path traversal allows to access any file on the server.
- 10.0CVE-2025-3322EPSS 0.7%
An improper neutralization of inputs used in expression language allows remote code execution with the highest privileges on the server.
- 9.4CVE-2025-3321EPSS 0.2%
A predefined administrative account is not documented and cannot be deactivated. This account cannot be misused from the network, only by local users on the server.
- 7.2CVE-2023-0888EPSS 0.6%
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An aut…
- 3.3CVE-2020-25168EPSS 0.2%
Hard-coded credentials in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 enable attackers with command line access to access the device’s Wi-Fi module.
- 7.1CVE-2020-25166EPSS 0.5%
An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware…
- 7.5CVE-2020-25164EPSS 0.6%
A vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to recover user credentials of the administrative interface.
- 7.5CVE-2020-25162EPSS 1.9%
A XPath injection vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows unauthenticated remote attackers to access sensitive information and escalate pr…
- 6.3CVE-2020-25160EPSS 0.2%
Improper access controls in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 enables attackers to extract and tamper with the devices network configuration.
- 6.1CVE-2020-25158EPSS 0.8%
A reflected cross-site scripting (XSS) vulnerability in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to inject arbitrary web script or HTML int…
- 7.2CVE-2020-25156EPSS 1.2%
Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.
- 6.1CVE-2020-25154EPSS 0.7%
An open redirect vulnerability in the administrative interface of the B. Braun Melsungen AG SpaceCom device Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to redirect users to malicious w…
- 8.1CVE-2020-25152EPSS 1.3%
A session fixation vulnerability in the B. Braun Melsungen AG SpaceCom administrative interface Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to hijack web sessions and escalate p…
- 8.8CVE-2020-25150EPSS 1.5%
A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with service user privileges to upload arbitrary files. By uploading …
- 6.7CVE-2020-16238EPSS 0.2%
A vulnerability in the configuration import mechanism of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with command line access to the underlying Linux…
- 8.8CVE-2021-33886EPSS 0.8%
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statement…
- 9.8CVE-2021-33885EPSS 5.6%
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This resul…
- 9.1CVE-2021-33884EPSS 1.0%
An Unrestricted Upload of File with Dangerous Type vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows remote attackers to upload any files to the /tmp directory of the device through the webpage API. This can result in critical…
- 7.5CVE-2021-33883EPSS 0.8%
A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to obtain sensitive information by snooping on the network traffic. The exposed data includes critical value…
- 8.6CVE-2021-33882EPSS 1.1%
A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking co…
- 7.8CVE-2020-25174EPSS 0.5%
A DLL hijacking vulnerability in the B. Braun OnlineSuite Version AP 3.0 and earlier allows local attackers to execute code on the system as a high privileged user.
- 9.8CVE-2020-25172EPSS 2.0%
A relative path traversal attack in the B. Braun OnlineSuite Version AP 3.0 and earlier allows unauthenticated attackers to upload or download arbitrary files.
- 7.8CVE-2020-25170EPSS 1.0%
An Excel Macro Injection vulnerability exists in the export feature in the B. Braun OnlineSuite Version AP 3.0 and earlier via multiple input fields that are mishandled in an Excel export.
- 6.1CVE-2017-6018EPSS 1.0%
An open redirect issue was discovered in B. Braun Medical SpaceCom module, which is integrated into the SpaceStation docking station: SpaceStation with SpaceCom module (integrated as part number 8713142U), software versions prior to Versio…