cve.bz

CVE-2021-33886 bbraun spacecom2

8.8 highExploit referenceEPSS 0.8%

High (8.8) in bbraun spacecom2. EPSS 0.8% estimated probability of exploitation activity in the next 30 days.

On this page
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.

Severity

CVSS 3.1

8.8 high
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
0.8% (unchanged from 0.8% a week ago; 0.8% 30 days ago) estimated probability of exploitation activity in the next 30 days; 58th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWE
CWE-134 Use of Externally-Controlled Format String
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2021-33886 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2021-33886?via=ask-ai-copy and https://cve.bz/cve/CVE-2021-33886?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2021-33886 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
bbraun:spacecom2before 012u000062

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-21-294-01B. Braun Infusomat Space Large Volume Pump2021-10-21Healthcare and Public HealthBattery Pack SP with Wi-Fi: All software Versions L81 and earlier that have been installed in a Perfusor Space Infusomat Space or Infusomat Space P pump, Data module compactPlus: All software Versions A10 and A11 that have been installed in a Perfusor compactPlus Infusomat compactPlus or Infusomat P compactPlus pump, SpaceStation with SpaceCom 2: All software Versions L81 and earlier and 2 morenone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-67 String Format Overflow in syslog() Likelihood: High · Severity: Very High
  • CAPEC-135 Format String Injection Likelihood: High · Severity: High

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON