- 3.5CVE-2023-30565EPSS 0.1%
An insecure connection between Systems Manager and CQI Reporter application could expose infusion data to an attacker.
- 6.9CVE-2023-30564EPSS 0.3%
Alaris Systems Manager does not perform input validation during the Device Import Function.
- 8.2CVE-2023-30563EPSS 0.4%
A malicious file could be uploaded into a System Manager User Import Function resulting in a hijacked session.
- 6.7CVE-2023-30562EPSS 0.2%
A GRE dataset file within Systems Manager can be tampered with and distributed to PCUs.
- 6.1CVE-2023-30561EPSS 0.2%
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
- 6.8CVE-2023-30560EPSS 0.3%
The configuration from the PCU can be modified without authentication using physical connection to the PCU.
- 5.7CVE-2023-30559EPSS 0.2%
The firmware update package for the wireless card is not properly signed and can be modified.