becton dickinson co vulnerabilities, newest first
Filter
By year: all 2026 2025 2024 2023 2022 2021
- 8.0CVE-2024-10476EPSS 0.2%
Default credentials are used in the above listed BD Diagnostic Solutions products. If exploited, threat actors may be able to access, modify or delete data, including sensitive information such as protected health information (PHI) and per…
- 3.5CVE-2023-30565EPSS 0.1%
An insecure connection between Systems Manager and CQI Reporter application could expose infusion data to an attacker.
- 6.9CVE-2023-30564EPSS 0.3%
Alaris Systems Manager does not perform input validation during the Device Import Function.
- 8.2CVE-2023-30563EPSS 0.4%
A malicious file could be uploaded into a System Manager User Import Function resulting in a hijacked session.
- 6.7CVE-2023-30562EPSS 0.2%
A GRE dataset file within Systems Manager can be tampered with and distributed to PCUs.
- 6.1CVE-2023-30561EPSS 0.2%
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
- 6.8CVE-2023-30560EPSS 0.3%
The configuration from the PCU can be modified without authentication using physical connection to the PCU.
- 5.7CVE-2023-30559EPSS 0.2%
The firmware update package for the wireless card is not properly signed and can be modified.