cve.bz

CVE-2025-14096

8.4 highEPSS 0.1%

High (8.4). EPSS 0.1% chance of exploitation in the next 30 days.

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Severity

CVSS 3.1

8.4 high
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
0.1% chance of exploitation in the next 30 days; higher than 2% of CVEs (scores of 2026-10-31)
CWEs
CWE-250, CWE-798
Published
Modified
Record from
the NVD, synced

References

Check your stackJSON