cve.bz

CVE-2024-9832 baxter life2000 ventilation system

9.3 criticalEPSS 0.2%

Critical (9.3) in baxter life2000 ventilation system. EPSS 0.2% estimated probability of exploitation activity in the next 30 days.

On this page
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

Severity

CVSS 3.1

9.3 critical
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS
0.2% (unchanged from 0.2% a week ago; 0.2% 30 days ago) estimated probability of exploitation activity in the next 30 days; 7th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: no · Technical impact: total, assessed 2024-11-18. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-307 Improper Restriction of Excessive Authentication Attempts
OWASP Top 10
A07:2025 Authentication Failures
Published
Modified
NVD analysis
Deferred: the NVD does not plan to analyze it
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2024-9832 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2024-9832?via=ask-ai-copy and https://cve.bz/cve/CVE-2024-9832?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2024-9832 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Baxter Life2000 Ventilation System06.08.00.00 and prior

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-24-319-01Baxter Life2000 Ventilation System2024-11-14Healthcare and Public HealthBaxter Life2000 Ventilation System: <=06.08.00.00none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-16 Dictionary-based Password Attack Likelihood: Medium · Severity: High
  • CAPEC-49 Password Brute Forcing Likelihood: Medium · Severity: High
  • CAPEC-560 Use of Known Domain Credentials Likelihood: High · Severity: High
  • CAPEC-565 Password Spraying Likelihood: High · Severity: High
  • CAPEC-600 Credential Stuffing Likelihood: High · Severity: High
  • CAPEC-652 Use of Known Kerberos Credentials Likelihood: Medium · Severity: High
  • CAPEC-653 Use of Known Operating System Credentials Likelihood: High · Severity: High

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON