{"cve":"CVE-2024-9832","nvd":{"published":"2024-11-14T21:15:22.593","modified":"2026-06-17T08:25:21.120","status":"Deferred","description":"There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.","cvss_v4":null,"cvss_v3":{"version":"3.1","vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","score":9.3},"cwes":["CWE-307"],"references":["https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"],"products":[],"source":"nvd","synced_at":"2026-10-19T21:57:07.693694550+00:00"},"epss":{"cve":"CVE-2024-9832","score":0.00176,"percentile":0.07138,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}