cve.bz

CVE-2024-48966 baxter life2000 ventilation system

10.0 criticalEPSS 0.7%

Critical (10.0) in baxter life2000 ventilation system. EPSS 0.7% estimated probability of exploitation activity in the next 30 days.

On this page
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Severity

CVSS 3.1

10.0 critical
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS
0.7% (unchanged from 0.7% a week ago; 0.7% 30 days ago) estimated probability of exploitation activity in the next 30 days; 54th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: yes · Technical impact: total, assessed 2024-11-15. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-306 Missing Authentication for Critical Functionality
OWASP Top 10
A07:2025 Authentication Failures
Published
Modified
NVD analysis
Deferred: the NVD does not plan to analyze it
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2024-48966 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2024-48966?via=ask-ai-copy and https://cve.bz/cve/CVE-2024-48966?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2024-48966 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Baxter Life2000 Ventilation System06.08.00.00 and prior

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-24-319-01Baxter Life2000 Ventilation System2024-11-14Healthcare and Public HealthBaxter Life2000 Ventilation System: <=06.08.00.00none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-12 Choosing Message Identifier Likelihood: High · Severity: High
  • CAPEC-36 Using Unpublished Interfaces or Functionality Likelihood: Medium · Severity: High
  • CAPEC-62 Cross Site Request Forgery Likelihood: High · Severity: Very High
  • CAPEC-166 Force the System to Reset Values Severity: Medium
  • CAPEC-216 Communication Channel Manipulation

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.