{"cve":"CVE-2024-48966","nvd":{"published":"2024-11-14T22:15:17.727","modified":"2026-06-17T07:59:08.223","status":"Deferred","description":"The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.","cvss_v4":null,"cvss_v3":{"version":"3.1","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","score":10.0},"cwes":["CWE-306"],"references":["https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"],"products":[],"source":"nvd","synced_at":"2026-10-19T21:57:07.693694550+00:00"},"epss":{"cve":"CVE-2024-48966","score":0.00695,"percentile":0.53016,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}