Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.

Severity

CVSS 3.1

7.5 high
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS
0.5% (unchanged from 0.5% a week ago; 0.5% 30 days ago) estimated probability of exploitation activity in the next 30 days; 40th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWE
CWE-319 Cleartext Transmission of Sensitive Information
OWASP Top 10
A04:2025 Cryptographic Failures
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2020-12048 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2020-12048?via=ask-ai-copy and https://cve.bz/cve/CVE-2020-12048?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2020-12048 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
baxter:phoenix_x36_firmware3.36
baxter:phoenix_x36_firmware3.40

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-20-170-03Baxter Phoenix Hemodialysis Delivery System (Update A)2020-06-18Healthcare and Public HealthPhoenix Hemodialysis Delivery System :SW 3.36 and 3.40none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-65 Sniff Application Code Likelihood: Low · Severity: High
  • CAPEC-102 Session Sidejacking Likelihood: High · Severity: High
  • CAPEC-117 Interception Likelihood: Low · Severity: Medium
  • CAPEC-383 Harvesting Information via API Event Monitoring Severity: Low
  • CAPEC-477 Signature Spoofing by Mixing Signed and Unsigned Content Likelihood: Low · Severity: High