{"cve":"CVE-2020-12048","nvd":{"published":"2020-06-29T14:15:11.990","modified":"2026-06-17T02:51:20.540","status":"Modified","description":"Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.","cvss_v4":null,"cvss_v3":{"version":"3.1","vector":"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N","score":7.5},"cwes":["CWE-319"],"references":["https://www.us-cert.gov/ics/advisories/icsma-20-170-03","https://www.us-cert.gov/ics/advisories/icsma-20-170-03"],"products":["baxter:phoenix_x36_firmware"],"source":"nvd","synced_at":"2026-10-19T21:06:22.076270124+00:00"},"epss":{"cve":"CVE-2020-12048","score":0.00452,"percentile":0.39292,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}