cve.bz

CVE-2018-8853 philips: Improper Privilege Management

8.8 highEPSS 0.4%

High (8.8) in philips. EPSS 0.4% estimated probability of exploitation activity in the next 30 days.

On this page
Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges in Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system.

Severity

CVSS 3.0

8.8 high
CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H
EPSS
0.4% (unchanged from 0.4% a week ago; 0.4% 30 days ago) estimated probability of exploitation activity in the next 30 days; 30th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-269 Improper Privilege Management
CWE-250 Execution with Unnecessary Privileges
OWASP Top 10
A06:2025 Insecure Design
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Fix and patched versions

Ask an AI about this CVE

Show the promptExplain CVE-2018-8853 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2018-8853?via=ask-ai-copy and https://cve.bz/cve/CVE-2018-8853?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2018-8853 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
philips:brilliance_firmware_64through 2.6.2
philips:brilliance_ict_sp_firmwarethrough 3.2.4
philips:brilliance_ict_firmwarethrough 4.1.6
philips:_brilliance_ct_big_bore_firmwarethrough 2.3.5

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-18-123-01Philips Brilliance Computed Tomography (CT) System (Update A)2018-05-03Healthcare and public HealthBrilliance CT Big Bore: Versions 2.3.5 and below, Brilliance iCT: Versions 4.1.6 and below, Brilliance iCT: SP Versions 3.2.4 and below and 1 morenone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-58 Restful Privilege Elevation Likelihood: High · Severity: High
  • CAPEC-122 Privilege Abuse Likelihood: High · Severity: Medium
  • CAPEC-233 Privilege Escalation
  • CAPEC-69 Target Programs with Elevated Privileges Likelihood: High · Severity: Very High
  • CAPEC-104 Cross Zone Scripting Likelihood: Medium · Severity: High
  • CAPEC-470 Expanding Control over the Operating System from the Database Severity: Very High

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON