In the default configuration, the archive accepts DICOM associations from an unregistered calling AE. A remote peer can use C-FIND and C-GET to enumerate and retrieve patient information and medical images. It can also use C-STORE to file fabricated study content under an existing patient identity.
- CVSS 4.0
- 9.3
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N - CVSS 3.1
- 9.1
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N - EPSS
- 0.3% chance of exploitation in the next 30 days; higher than 19% of CVEs (scores of 2026-10-25)
- CWE
- CWE-306