cve.bz

CVE-2025-4394 medtronic

6.8 mediumEPSS 0.2%

Medium (6.8) in medtronic. EPSS 0.2% estimated probability of exploitation activity in the next 30 days.

On this page
Medtronic MyCareLink Patient Monitor uses an unencrypted filesystem on internal storage, which allows an attacker with physical access to read and modify files. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025

Severity

CVSS 3.1

6.8 medium
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
0.2% (unchanged from 0.2% a week ago; 0.2% 30 days ago) estimated probability of exploitation activity in the next 30 days; 12th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: no · Technical impact: total, assessed 2025-07-24. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-312 Cleartext Storage of Sensitive Information
OWASP Top 10
A06:2025 Insecure Design
Published
Modified
NVD analysis
Deferred: the NVD does not plan to analyze it
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2025-4394 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2025-4394?via=ask-ai-copy and https://cve.bz/cve/CVE-2025-4394?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2025-4394 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Medtronic MyCareLink Patient Monitor 24950before June 25, 2025
Medtronic MyCareLink Patient Monitor 24952before June 25, 2025

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-25-205-01Medtronic MyCareLink Patient Monitor (Update A)2025-07-24Healthcare and Public HealthMedtronic MyCareLink Patient Monitor model 24950: vers:all/*, Medtronic MyCareLink Patient Monitor model 24952: vers:all/*none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-37 Retrieve Embedded Sensitive Data Likelihood: High · Severity: Very High