cve.bz

CVE-2025-14095 radiometer medical aps: Improper Access Control

6.8 mediumEPSS 0.2%

Medium (6.8) in radiometer medical aps. EPSS 0.2% estimated probability of exploitation activity in the next 30 days.

On this page
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.  Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

Severity

CVSS 3.1

6.8 medium
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
0.2% (unchanged from 0.2% a week ago; 0.2% 30 days ago) estimated probability of exploitation activity in the next 30 days; 7th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: no · Technical impact: partial, assessed 2025-12-17. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWEs
CWE-284 Improper Access Control
CWE-693 Protection Mechanism Failure
OWASP Top 10
A01:2025 Broken Access Control, A06:2025 Insecure Design
Published
Modified
NVD analysis
Deferred: the NVD does not plan to analyze it
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2025-14095 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2025-14095?via=ask-ai-copy and https://cve.bz/cve/CVE-2025-14095?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2025-14095 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS AnalyzersAll application software versions with Windows 7, Windows XP as
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS AnalyzersApplication software versions < 3.5MR11 with Windows 10 as under
Radiometer Medical Aps AQT90 FLEX AnalyzersAll Application software versions <= 8.13 MR2
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX AnalyzersApplication software versions < 6.20 MR2 with Windows 7, Windows
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX AnalyzersApplication software versions < 6.20 MR2 with Windows 10 as unde
Radiometer Medical Aps ABL9 AnalyzersApplication software versions < 1.5.0