cve.bz

CVE-2024-9834 baxter life2000 ventilation system

9.3 criticalEPSS 0.1%

Critical (9.3) in baxter life2000 ventilation system. EPSS 0.1% estimated probability of exploitation activity in the next 30 days.

On this page
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Severity

CVSS 3.1

9.3 critical
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS
0.1% (unchanged from 0.1% a week ago; 0.1% 30 days ago) estimated probability of exploitation activity in the next 30 days; 3rd percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: no · Technical impact: total, assessed 2024-11-15. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-319 Cleartext Transmission of Sensitive Information
OWASP Top 10
A04:2025 Cryptographic Failures
Published
Modified
NVD analysis
Deferred: the NVD does not plan to analyze it
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2024-9834 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2024-9834?via=ask-ai-copy and https://cve.bz/cve/CVE-2024-9834?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2024-9834 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Baxter Life2000 Ventilation System06.08.00.00 and prior

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-24-319-01Baxter Life2000 Ventilation System2024-11-14Healthcare and Public HealthBaxter Life2000 Ventilation System: <=06.08.00.00none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-65 Sniff Application Code Likelihood: Low · Severity: High
  • CAPEC-102 Session Sidejacking Likelihood: High · Severity: High
  • CAPEC-117 Interception Likelihood: Low · Severity: Medium
  • CAPEC-383 Harvesting Information via API Event Monitoring Severity: Low
  • CAPEC-477 Signature Spoofing by Mixing Signed and Unsigned Content Likelihood: Low · Severity: High

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.