CVE-2024-48970 baxter life2000 ventilation system

9.3 criticalEPSS 0.2%

Critical (9.3) in baxter life2000 ventilation system. EPSS 0.2% estimated probability of exploitation activity in the next 30 days.

On this page
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.

Severity

CVSS 3.1

9.3 critical
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS
0.2% (unchanged from 0.2% a week ago; 0.2% 30 days ago) estimated probability of exploitation activity in the next 30 days; 12th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: no · Technical impact: total, assessed 2024-11-18. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-1191 On-Chip Debug and Test Interface With Improper Access Control
Published
Modified
NVD analysis
Deferred: the NVD does not plan to analyze it
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2024-48970 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2024-48970?via=ask-ai-copy and https://cve.bz/cve/CVE-2024-48970?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2024-48970 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Baxter Life2000 Ventilation System06.08.00.00 and prior

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-24-319-01Baxter Life2000 Ventilation System2024-11-14Healthcare and Public HealthBaxter Life2000 Ventilation System: <=06.08.00.00none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-1 Accessing Functionality Not Properly Constrained by ACLs Likelihood: High · Severity: High
  • CAPEC-180 Exploiting Incorrectly Configured Access Control Security Levels Likelihood: High · Severity: Medium

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON