cve.bz

CVE-2024-48967 baxter life2000 ventilation system: Insufficient Logging

10.0 criticalEPSS 0.6%

Critical (10.0) in baxter life2000 ventilation system. EPSS 0.6% estimated probability of exploitation activity in the next 30 days.

On this page
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.

Severity

CVSS 3.1

10.0 critical
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS
0.6% (unchanged from 0.6% a week ago; 0.6% 30 days ago) estimated probability of exploitation activity in the next 30 days; 51st percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: yes · Technical impact: total, assessed 2024-11-15. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-778 Insufficient Logging
OWASP Top 10
A09:2025 Logging & Alerting Failures
Published
Modified
NVD analysis
Deferred: the NVD does not plan to analyze it
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2024-48967 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2024-48967?via=ask-ai-copy and https://cve.bz/cve/CVE-2024-48967?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2024-48967 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Baxter Life2000 Ventilation System06.08.00.00 and prior

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-24-319-01Baxter Life2000 Ventilation System2024-11-14Healthcare and Public HealthBaxter Life2000 Ventilation System: <=06.08.00.00none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

References

Check your stackJSON