cve.bz

CVE-2022-4992 dräger

8.8 highEPSS 0.2%

High (8.8) in dräger. EPSS 0.2% estimated probability of exploitation activity in the next 30 days.

On this page
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.

Severity

CVSS 4.0

8.8 high
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N

CVSS 3.1

8.6 high
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
EPSS
0.2% (unchanged from 0.2% a week ago; 0.2% 30 days ago) estimated probability of exploitation activity in the next 30 days; 5th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: yes · Technical impact: partial, assessed 2026-06-03. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-345 Insufficient Verification of Data Authenticity
OWASP Top 10
A08:2025 Software or Data Integrity Failures
Published
Modified
NVD analysis
Awaiting analysis
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2022-4992 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2022-4992?via=ask-ai-copy and https://cve.bz/cve/CVE-2022-4992?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2022-4992 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Dräger Infinity Acute Care Systembefore VG4.2
Dräger Infinity Acute Care Systembefore VG4.1.1
Dräger Infinity Acute Care Systembefore VG4.0.3
Dräger Standalone Infinity M540 patient monitorbefore VG4.2
Dräger Standalone Infinity M540 patient monitorbefore VG4.1.1