cve.bz

CVE-2022-32537 medtronic: Protection Mechanism Failure

4.8 mediumEPSS 0.4%

Medium (4.8) in medtronic. EPSS 0.4% estimated probability of exploitation activity in the next 30 days.

On this page
A vulnerability exists which could allow an unauthorized user to learn aspects of the communication protocol used to pair system components while the pump is being paired with other system components. Exploitation requires nearby wireless signal proximity with the patient and the device; advanced technical knowledge is required for exploitation. Please refer to the Medtronic Product Security Bulletin for guidance

Severity

CVSS 3.1

4.8 medium
CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:U/C:N/I:H/A:N
EPSS
0.4% (unchanged from 0.4% a week ago; 0.3% 30 days ago) estimated probability of exploitation activity in the next 30 days; 30th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: no · Technical impact: partial, assessed 2025-04-29. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-693 Protection Mechanism Failure
OWASP Top 10
A06:2025 Insecure Design
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Fix and patched versions

Ask an AI about this CVE

Show the promptExplain CVE-2022-32537 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2022-32537?via=ask-ai-copy and https://cve.bz/cve/CVE-2022-32537?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2022-32537 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Medtronic Minimed 600 Series Insulin Pump620G, 630G, 640G, 670G

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-22-263-01Medtronic NGP 600 Series Insulin Pumps2022-09-20Healthcare and Public HealthMiniMed 620G: MMT-1710, MiniMed 630G: MMT-1715 MMT-1754 MMT-1755, MiniMed 640G: MMT-1711 MMT-1712 MMT-1751 MMT-1752 and 1 morenone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits. More on the page of CWE-693 (17).

References

Check your stackJSON