cve.bz

CVE-2022-3027 contechealth cms8000 firmware: Improper Access Control

5.7 mediumEPSS 0.3%

Medium (5.7) in contechealth cms8000 firmware. EPSS 0.3% estimated probability of exploitation activity in the next 30 days.

On this page
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.

Severity

CVSS 3.1

5.7 medium
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:H/A:N
EPSS
0.3% (unchanged from 0.3% a week ago; 0.3% 30 days ago) estimated probability of exploitation activity in the next 30 days; 26th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: no · Technical impact: partial, assessed 2025-04-16. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-284 Improper Access Control
OWASP Top 10
A01:2025 Broken Access Control
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2022-3027 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2022-3027?via=ask-ai-copy and https://cve.bz/cve/CVE-2022-3027?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2022-3027 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient MonitorAll

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-22-244-01Contec Health CMS8000 Patient Monitor (Update A)2022-09-01Healthcare and Public HealthContec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor: <=smart3250-2.6.27-wlan2.1.7.cramfs, Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor: <=CMS7.820.075.08/0.740.75, Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor: <=CMS7.820.120.01/0.930.95none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits. More on the page of CWE-284 (17).

References

Check your stackJSON