cve.bz

CVE-2021-38392 bostonscientific zoom latitude pogrammer\/recorder\/monitor 3120 firmware: Improper Access Control

7.6 highEPSS 0.3%

High (7.6) in bostonscientific zoom latitude pogrammer\/recorder\/monitor 3120 firmware. EPSS 0.3% estimated probability of exploitation activity in the next 30 days.

On this page
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.

Severity

CVSS 3.1

7.6 high
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS
0.3% (unchanged from 0.3% a week ago; 0.3% 30 days ago) estimated probability of exploitation activity in the next 30 days; 18th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWE
CWE-284 Improper Access Control
OWASP Top 10
A01:2025 Broken Access Control
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2021-38392 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2021-38392?via=ask-ai-copy and https://cve.bz/cve/CVE-2021-38392?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2021-38392 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

  • bostonscientific:zoom_latitude_pogrammer\/recorder\/monitor_3120_firmware

Versions

From the CNA's record in the CVE List.

ProductAffected
Boston Scientific ZOOM LATITUDEModel 3120

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-21-273-01Boston Scientific Zoom Latitude2021-09-30Healthcare and Public HealthZOOM LATITUDE Programmer/Recorder/Monitor Model 3120 - Boston Scientific reports these vulnerabilities affects the ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) Model 3120none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits. More on the page of CWE-284 (17).

References

Check your stackJSON