cve.bz

CVE-2021-37162

9.8 criticalEPSS 3.3%

Critical (9.8) in swisslog-healthcare hmi-3 control panel firmware. EPSS 3.3% chance of exploitation in the next 30 days.

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.

Severity

CVSS 3.1

9.8 critical
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
3.3% chance of exploitation in the next 30 days; higher than 88% of CVEs (scores of 2026-10-31)
CWE
CWE-120
Published
Modified
Record from
the NVD, synced

Affected

References