CVE-2020-25183 medtronic mycarelink smart model 25000 firmware: Improper Authentication

8.8 highEPSS 0.8%

High (8.8) in medtronic mycarelink smart model 25000 firmware. EPSS 0.8% estimated probability of exploitation activity in the next 30 days.

On this page
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

Severity

CVSS 3.1

8.8 high
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
0.8% (unchanged from 0.8% a week ago; 0.8% 30 days ago) estimated probability of exploitation activity in the next 30 days; 57th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWE
CWE-287 Improper Authentication
OWASP Top 10
A07:2025 Authentication Failures
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2020-25183 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2020-25183?via=ask-ai-copy and https://cve.bz/cve/CVE-2020-25183?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2020-25183 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Medtronic Smart Model 25000 Patient ReaderAll versions

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-20-345-01Medtronic MyCareLink Smart2020-12-08Healthcare and Public HealthSmart Model 25000 Patient Reader: all versionsnone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-22 Exploiting Trust in Client Likelihood: High · Severity: High
  • CAPEC-57 Utilizing REST's Trust in the System Resource to Obtain Sensitive Data Likelihood: Medium · Severity: Very High
  • CAPEC-94 Adversary in the Middle (AiTM) Likelihood: High · Severity: Very High
  • CAPEC-114 Authentication Abuse Severity: Medium
  • CAPEC-115 Authentication Bypass Severity: Medium
  • CAPEC-151 Identity Spoofing Likelihood: Medium · Severity: Medium
  • CAPEC-194 Fake the Source of Data Severity: Medium
  • CAPEC-593 Session Hijacking Likelihood: High · Severity: Very High
  • CAPEC-633 Token Impersonation Severity: Medium
  • CAPEC-650 Upload a Web Shell to a Web Server Severity: High

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON