Medium (6.5) in philips patient information center ix. EPSS 0.6% chance of exploitation in the next 30 days.
In Patient Information Center iX (PICiX) Versions C.02, C.03, the
software parses a formatted message or structure but does not handle or
incorrectly handles a length field that is inconsistent with the actual
length of the associated data, causing the application on the
surveillance station to restart.
Severity
CVSS 3.1
6.5 medium
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
EPSS
0.6% chance of exploitation in the next 30 days; higher than 48% of CVEs (scores of 2026-10-31)