CVE-2020-15486

6.5 mediumEPSS 0.5%

Medium (6.5) in drtrust electrocardiogram pen firmware. EPSS 0.5% chance of exploitation in the next 30 days.

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Severity

CVSS 3.1

6.5 medium
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS
0.5% chance of exploitation in the next 30 days; higher than 43% of CVEs (scores of 2026-10-31)