cve.bz

CVE-2020-12043 baxter sigma spectrum infusion system firmware

9.8 criticalEPSS 2.2%

Critical (9.8) in baxter sigma spectrum infusion system firmware. EPSS 2.2% estimated probability of exploitation activity in the next 30 days.

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted.

Severity

CVSS 3.1

9.8 critical
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
2.2% (unchanged from 2.2% a week ago; 2.1% 30 days ago) estimated probability of exploitation activity in the next 30 days; 83rd percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWE
CWE-672 Operation on a Resource after Expiration or Release
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2020-12043 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2020-12043?via=ask-ai-copy and https://cve.bz/cve/CVE-2020-12043?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2020-12043 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
baxter:sigma_spectrum_infusion_system_firmware8.0

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-20-170-04Baxter Sigma Spectrum Infusion Pumps (Update B)2020-06-18Healthcare and Public HealthBaxter Spectrum: v8.x model 35700BAX2, Sigma Spectrum LVP: v6.x with Wireless Battery Modules v9 v11 v13 v14 v15 v16 v16D38 v17 v17D19 v20D29 to v20D32 and v22D24 to v22D28, Sigma Spectrum: v6.x model 35700BAX and 3 morenone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

References

Check your stackJSON