cve.bz

CVE-2020-12039 baxter sigma spectrum infusion system firmware: Use of Hard-coded Credentials

2.4 lowEPSS 0.3%

Low (2.4) in baxter sigma spectrum infusion system firmware. EPSS 0.3% estimated probability of exploitation activity in the next 30 days.

On this page
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.

Severity

CVSS 3.1

2.4 low
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
EPSS
0.3% (unchanged from 0.3% a week ago; 0.3% 30 days ago) estimated probability of exploitation activity in the next 30 days; 29th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-798 Use of Hard-coded Credentials
CWE-259 Use of Hard-coded Password
OWASP Top 10
A07:2025 Authentication Failures
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2020-12039 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2020-12039?via=ask-ai-copy and https://cve.bz/cve/CVE-2020-12039?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2020-12039 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
baxter:sigma_spectrum_infusion_system_firmwarefrom 6.0 through 6.05
baxter:sigma_spectrum_infusion_system_firmware8.0

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-20-170-04Baxter Sigma Spectrum Infusion Pumps (Update B)2020-06-18Healthcare and Public HealthBaxter Spectrum: v8.x model 35700BAX2, Sigma Spectrum LVP: v6.x with Wireless Battery Modules v9 v11 v13 v14 v15 v16 v16D38 v17 v17D19 v20D29 to v20D32 and v22D24 to v22D28, Sigma Spectrum: v6.x model 35700BAX and 3 morenone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-70 Try Common or Default Usernames and Passwords Likelihood: Medium · Severity: High
  • CAPEC-191 Read Sensitive Constants Within an Executable Severity: Low

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON