cve.bz

CVE-2019-6538 medtronic: Missing Authorization

6.5 mediumEPSS 0.9%

Medium (6.5) in medtronic. EPSS 0.9% estimated probability of exploitation activity in the next 30 days.

On this page
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

Severity

CVSS 3.1

6.5 medium
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
EPSS
0.9% (unchanged from 0.9% a week ago; 0.9% 30 days ago) estimated probability of exploitation activity in the next 30 days; 60th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-306 Missing Authentication for Critical Functionality
CWE-862 Missing Authorization
CWE-284 Improper Access Control
OWASP Top 10
A01:2025 Broken Access Control, A07:2025 Authentication Failures
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2019-6538 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2019-6538?via=ask-ai-copy and https://cve.bz/cve/CVE-2019-6538?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2019-6538 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
medtronic:mycarelink_monitor_firmware24950
medtronic:mycarelink_monitor_firmware24952
medtronic:carelink_monitor_firmware2490c

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-19-080-01Medtronic Conexus Radio Frequency Telemetry Protocol (Update C)2019-03-21Healthcare and Public HealthMyCareLink Monitor: Versions 24950 and 24952, Brava CRT-D: (all models), Amplia CRT-D: (all models) and 19 morenone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-665 Exploitation of Thunderbolt Protection Flaws Likelihood: Low · Severity: Very High
  • CAPEC-12 Choosing Message Identifier Likelihood: High · Severity: High
  • CAPEC-36 Using Unpublished Interfaces or Functionality Likelihood: Medium · Severity: High
  • CAPEC-62 Cross Site Request Forgery Likelihood: High · Severity: Very High
  • CAPEC-166 Force the System to Reset Values Severity: Medium
  • CAPEC-216 Communication Channel Manipulation

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits. More on the page of CWE-284 (17).

References

Check your stackJSON