cve.bz

CVE-2019-25720 dräger

7.1 highEPSS 0.2%

High (7.1) in dräger. EPSS 0.2% estimated probability of exploitation activity in the next 30 days.

On this page
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

Severity

CVSS 4.0

7.1 high
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N

CVSS 3.1

6.5 medium
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
EPSS
0.2% (unchanged from 0.2% a week ago; 0.2% 30 days ago) estimated probability of exploitation activity in the next 30 days; 10th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CISA's assessment (SSVC)
Exploitation: none · Automatable: no · Technical impact: partial, assessed 2026-06-03. From CISA's Vulnrichment in the CVE List: whether it is exploited, whether exploiting it can be automated, and how much control it gives.
CWE
CWE-1286 Improper Validation of Syntactic Correctness of Input
Published
Modified
NVD analysis
Awaiting analysis
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2019-25720 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2019-25720?via=ask-ai-copy and https://cve.bz/cve/CVE-2019-25720?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2019-25720 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Dräger SC 6002XLSC 6002XL
Dräger SC6802XLSC6802XL
Dräger SC 7000SC 7000
Dräger SC8000SC8000
Dräger SC90000 XLSC90000 XL

Attack patterns (CAPEC)

  • CAPEC-66 SQL Injection Likelihood: High · Severity: High
  • CAPEC-676 NoSQL Injection Likelihood: High · Severity: High

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON