cve.bz

CVE-2019-25719

8.8 highEPSS 0.1%

High (8.8). EPSS 0.1% chance of exploitation in the next 30 days.

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Severity

CVSS 4.0

8.8 high
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N

CVSS 3.1

8.6 high
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
EPSS
0.1% chance of exploitation in the next 30 days; higher than 2% of CVEs (scores of 2026-10-31)
CWE
CWE-924
Published
Modified
Record from
the NVD, synced

References

Check your stackJSON