cve.bz

CVE-2019-18248

4.3 mediumEPSS 0.4%

Medium (4.3) in biotronik. EPSS 0.4% chance of exploitation in the next 30 days.

BIOTRONIK CardioMessenger II, The affected products transmit credentials in clear-text prior to switching to an encrypted communication channel. An attacker can disclose the product’s client credentials for connecting to the BIOTRONIK Remote Communication infrastructure.

Severity

CVSS 3.1

4.3 medium
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
EPSS
0.4% chance of exploitation in the next 30 days; higher than 30% of CVEs (scores of 2026-10-31)
CWE
CWE-319
Published
Modified
Record from
the NVD, synced

Affected

References

Check your stackJSON