cve.bz

CVE-2018-8868 medtronic: Exposed Dangerous Method or Function

6.2 mediumEPSS 0.3%

Medium (6.2) in medtronic. EPSS 0.3% estimated probability of exploitation activity in the next 30 days.

On this page
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Severity

CVSS 3.1

6.2 medium
CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
EPSS
0.3% (unchanged from 0.3% a week ago; 0.3% 30 days ago) estimated probability of exploitation activity in the next 30 days; 27th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWE
CWE-749 Exposed Dangerous Method or Function
OWASP Top 10
A01:2025 Broken Access Control
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2018-8868 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2018-8868?via=ask-ai-copy and https://cve.bz/cve/CVE-2018-8868?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2018-8868 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Medtronic 24950 MyCareLink MonitorAll versions
Medtronic 24952 MyCareLink MonitorAll versions

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-18-179-01Medtronic MyCareLink Patient Monitor2018-06-28Healthcare and Public Health24952 MyCareLink Monitor: all versions, 24950 MyCareLink Monitor: all versionsnone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References