cve.bz

CVE-2018-8861 philips: Exposure of Resource to Wrong Sphere

8.7 highEPSS 0.4%

High (8.7) in philips. EPSS 0.4% estimated probability of exploitation activity in the next 30 days.

On this page
Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system.

Severity

CVSS 3.0

8.7 high
CVSS:3.0/AV:L/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:L
EPSS
0.4% (unchanged from 0.4% a week ago; 0.4% 30 days ago) estimated probability of exploitation activity in the next 30 days; 35th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWE
CWE-668 Exposure of Resource to Wrong Sphere
OWASP Top 10
A01:2025 Broken Access Control
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Fix and patched versions

Ask an AI about this CVE

Show the promptExplain CVE-2018-8861 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2018-8861?via=ask-ai-copy and https://cve.bz/cve/CVE-2018-8861?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2018-8861 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
philips:brilliance_firmware_64through 2.6.2
philips:brilliance_ict_sp_firmwarethrough 3.2.4
philips:brilliance_ict_firmwarethrough 4.1.6
philips:_brilliance_ct_big_bore_firmwarethrough 2.3.5

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-18-123-01Philips Brilliance Computed Tomography (CT) System (Update A)2018-05-03Healthcare and public HealthBrilliance CT Big Bore: Versions 2.3.5 and below, Brilliance iCT: Versions 4.1.6 and below, Brilliance iCT: SP Versions 3.2.4 and below and 1 morenone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.