CVE-2018-14801 philips: Use of Hard-coded Credentials

6.2 mediumEPSS 0.4%

Medium (6.2) in philips. EPSS 0.4% estimated probability of exploitation activity in the next 30 days.

On this page
In Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs, all versions prior to May 2018, an attacker with both the superuser password and physical access can enter the superuser password that can be used to access and modify all settings on the device, as well as allow the user to reset existing passwords.

Severity

CVSS 3.0

6.2 medium
CVSS:3.0/AV:P/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
EPSS
0.4% (unchanged from 0.4% a week ago; 0.4% 30 days ago) estimated probability of exploitation activity in the next 30 days; 36th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWE
CWE-798 Use of Hard-coded Credentials
OWASP Top 10
A07:2025 Authentication Failures
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Fix and patched versions

Ask an AI about this CVE

Show the promptExplain CVE-2018-14801 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2018-14801?via=ask-ai-copy and https://cve.bz/cve/CVE-2018-14801?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2018-14801 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Philips PageWriter TC10, TC20, TC30, TC50, TC70 CardiographsAll versions prior to May 2018.

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-18-228-01Philips PageWriter TC10, TC20, TC30, TC50, and TC70 Cardiographs (Update A)2018-08-16Healthcare and Public HealthPageWriter TC10 TC20 TC30 TC50 TC70 Cardiographs: All versions prior to May 2018none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-70 Try Common or Default Usernames and Passwords Likelihood: Medium · Severity: High
  • CAPEC-191 Read Sensitive Constants Within an Executable Severity: Low

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.