High (7.8) in philips. EPSS 0.4% chance of exploitation in the next 30 days.
In Philips' IntelliSpace Cardiovascular (ISCV) products (ISCV Version 2.x or prior and Xcelera Version 4.1 or prior), an attacker with escalated privileges could access folders which contain executables where authenticated users have write permissions, and could then execute arbitrary code with local administrative permissions.
Severity
CVSS 3.1
7.8 high
CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
EPSS
0.4% chance of exploitation in the next 30 days; higher than 33% of CVEs (scores of 2026-10-31)