CVE-2018-10622 medtronic: Insufficiently Protected Credentials

5.2 mediumEPSS 0.4%

Medium (5.2) in medtronic. EPSS 0.4% estimated probability of exploitation activity in the next 30 days.

On this page
Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.

Severity

CVSS 3.1

5.2 medium
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:N
EPSS
0.4% (unchanged from 0.4% a week ago; 0.4% 30 days ago) estimated probability of exploitation activity in the next 30 days; 30th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-522 Insufficiently Protected Credentials
CWE-313 Cleartext Storage in a File or on Disk
OWASP Top 10
A06:2025 Insecure Design
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2018-10622 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2018-10622?via=ask-ai-copy and https://cve.bz/cve/CVE-2018-10622?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2018-10622 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Medtronic 24950 MyCareLink Monitorbefore February 25, 2026
Medtronic 24952 MyCareLink Monitorbefore February 25, 2026

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-25-205-01Medtronic MyCareLink Patient Monitor (Update A)2025-07-24Healthcare and Public HealthMedtronic MyCareLink Patient Monitor model 24950: vers:all/*, Medtronic MyCareLink Patient Monitor model 24952: vers:all/*none listed
ICSMA-18-219-01Medtronic MyCareLink 24950 Patient Monitor (Update A)2018-08-07Healthcare and Public HealthMedtronic 24950 MyCareLink Monitor: vers:all/*, Medtronic 24952 MyCareLink Monitor: vers:all/*none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits. More on the page of CWE-522 (13).

References