cve.bz

CVE-2018-10601 philips: Out-of-bounds Write

8.2 highEPSS 0.4%

High (8.2) in philips. EPSS 0.4% estimated probability of exploitation activity in the next 30 days.

On this page
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.

Severity

CVSS 3.1

8.2 high
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H
EPSS
0.4% (unchanged from 0.4% a week ago; 0.4% 30 days ago) estimated probability of exploitation activity in the next 30 days; 31st percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-787 Out-of-bounds Write
CWE-121 Stack-based Buffer Overflow
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2018-10601 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2018-10601?via=ask-ai-copy and https://cve.bz/cve/CVE-2018-10601?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2018-10601 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal MonitorsThe following IntelliVue Patient Monitors versions are affected:

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-18-156-01Philips' IntelliVue Patient and Avalon Fetal Monitors2018-06-05Healthcare and Public HealthAvalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50: with software Revisions F.0 G.0 and J.3, IntelliVue Patient Monitors MX (MX400-550): Rev J-M and (X3/MX100 for Rev M only), IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800): Rev B-Mnone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

References

Check your stackJSON