cve.bz

CVE-2018-10597 philips: Out-of-bounds Write

8.3 highEPSS 0.4%

High (8.3) in philips. EPSS 0.4% estimated probability of exploitation activity in the next 30 days.

On this page
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to access memory ("write-what-where") from an attacker-chosen device address within the same subnet.

Severity

CVSS 3.1

8.3 high
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS
0.4% (unchanged from 0.4% a week ago; 0.4% 30 days ago) estimated probability of exploitation activity in the next 30 days; 34th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-787 Out-of-bounds Write
CWE-287 Improper Authentication
OWASP Top 10
A07:2025 Authentication Failures
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2018-10597 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2018-10597?via=ask-ai-copy and https://cve.bz/cve/CVE-2018-10597?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2018-10597 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal MonitorsThe following IntelliVue Patient Monitors versions are affected:

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-18-156-01Philips' IntelliVue Patient and Avalon Fetal Monitors2018-06-05Healthcare and Public HealthAvalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50: with software Revisions F.0 G.0 and J.3, IntelliVue Patient Monitors MX (MX400-550): Rev J-M and (X3/MX100 for Rev M only), IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800): Rev B-Mnone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-22 Exploiting Trust in Client Likelihood: High · Severity: High
  • CAPEC-57 Utilizing REST's Trust in the System Resource to Obtain Sensitive Data Likelihood: Medium · Severity: Very High
  • CAPEC-94 Adversary in the Middle (AiTM) Likelihood: High · Severity: Very High
  • CAPEC-114 Authentication Abuse Severity: Medium
  • CAPEC-115 Authentication Bypass Severity: Medium
  • CAPEC-151 Identity Spoofing Likelihood: Medium · Severity: Medium
  • CAPEC-194 Fake the Source of Data Severity: Medium
  • CAPEC-593 Session Hijacking Likelihood: High · Severity: Very High
  • CAPEC-633 Token Impersonation Severity: Medium
  • CAPEC-650 Upload a Web Shell to a Web Server Severity: High

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON