cve.bz

CVE-2018-10596 medtronic 2090 carelink programmer firmware

7.1 highEPSS 1.3%

High (7.1) in medtronic 2090 carelink programmer firmware. EPSS 1.3% estimated probability of exploitation activity in the next 30 days.

On this page
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

Severity

CVSS 3.1

7.1 high
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
EPSS
1.3% (unchanged from 1.3% a week ago; 1.3% 30 days ago) estimated probability of exploitation activity in the next 30 days; 72nd percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-200 Exposure of Sensitive Information to an Unauthorized Actor
CWE-923 Improper Restriction of Communication Channel to Intended Endpoints
OWASP Top 10
A01:2025 Broken Access Control
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2018-10596 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2018-10596?via=ask-ai-copy and https://cve.bz/cve/CVE-2018-10596?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2018-10596 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the CNA's record in the CVE List.

ProductAffected
Medtronic 2090 CareLink ProgrammerAll versions
Medtronic 29901 Encore ProgrammerAll versions

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-18-058-01Medtronic 2090 Carelink Programmer Vulnerabilities (Update C)2018-02-27Healthcare and Public Health29901 Encore Programmer: all versions, 2090 CareLink Programmer: all versionsnone listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-161 Infrastructure Manipulation Severity: High
  • CAPEC-481 Contradictory Destinations in Traffic Routing Schemes Likelihood: Medium · Severity: High
  • CAPEC-501 Android Activity Hijack Severity: Medium
  • CAPEC-697 DHCP Spoofing Likelihood: Low · Severity: High

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits. More on the page of CWE-200 (59).

References

Check your stackJSON