CVE-2017-12725 smiths-medical medfusion 4000 wireless syringe infusion pump: Use of Hard-coded Credentials

5.6 mediumEPSS 0.8%

Medium (5.6) in smiths-medical medfusion 4000 wireless syringe infusion pump. EPSS 0.8% estimated probability of exploitation activity in the next 30 days.

On this page
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Severity

CVSS 3.0

5.6 medium
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L
EPSS
0.8% (unchanged from 0.8% a week ago; 0.8% 30 days ago) estimated probability of exploitation activity in the next 30 days; 59th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWE
CWE-798 Use of Hard-coded Credentials
OWASP Top 10
A07:2025 Authentication Failures
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2017-12725 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2017-12725?via=ask-ai-copy and https://cve.bz/cve/CVE-2017-12725?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2017-12725 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
smiths-medical:medfusion_4000_wireless_syringe_infusion_pump1.1
smiths-medical:medfusion_4000_wireless_syringe_infusion_pump1.5
smiths-medical:medfusion_4000_wireless_syringe_infusion_pump1.6

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSMA-17-250-02ASmiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities (Update A)2017-09-07Medfusion 4000 Wireless Syringe Infusion Pump: Version 1.1, Medfusion 4000 Wireless Syringe Infusion Pump: Version 1.6, Medfusion 4000 Wireless Syringe Infusion Pump: Version 1.5none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-70 Try Common or Default Usernames and Passwords Likelihood: Medium · Severity: High
  • CAPEC-191 Read Sensitive Constants Within an Executable Severity: Low

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.