cve.bz

CVE-2017-12718 smiths-medical medfusion 4000 wireless syringe infusion pump: Classic Buffer Overflow

8.1 highExploitEPSS 13%

High (8.1) in smiths-medical medfusion 4000 wireless syringe infusion pump. EPSS 13% estimated probability of exploitation activity in the next 30 days.

On this page
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

Severity

CVSS 3.0

8.1 high
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
13% (unchanged from 13% a week ago; 13% 30 days ago) estimated probability of exploitation activity in the next 30 days; 96th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-119 Improper Restriction of Operations within the Bounds of a Memory Buffer
CWE-120 Buffer Copy without Checking Size of Input (Classic Buffer Overflow)
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2017-12718 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2017-12718?via=ask-ai-copy and https://cve.bz/cve/CVE-2017-12718?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2017-12718 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
smiths-medical:medfusion_4000_wireless_syringe_infusion_pump1.1
smiths-medical:medfusion_4000_wireless_syringe_infusion_pump1.5
smiths-medical:medfusion_4000_wireless_syringe_infusion_pump1.6

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSA-17-285-04ANXP Semiconductors MQX RTOS (Update A)2017-10-12Communications, Critical Manufacturing, Healthcare and Public Health, and TransportationMQX RTOS: Version 4.1 and prior versions, MQX RTOS: Version 5.0 and prior versionsnone listed
ICSMA-17-250-02ASmiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities (Update A)2017-09-07Medfusion 4000 Wireless Syringe Infusion Pump: Version 1.1, Medfusion 4000 Wireless Syringe Infusion Pump: Version 1.6, Medfusion 4000 Wireless Syringe Infusion Pump: Version 1.5none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Public exploits

Attack patterns (CAPEC)

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits. More on the page of CWE-119 (12), CWE-120 (13).

References

Check your stackJSON