cve.bz

CVE-2017-12716

6.5 mediumEPSS 0.2%

Medium (6.5) in abbott. EPSS 0.2% chance of exploitation in the next 30 days.

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

Severity

CVSS 3.0

6.5 medium
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS
0.2% chance of exploitation in the next 30 days; higher than 15% of CVEs (scores of 2026-10-31)
CWEs
CWE-319, CWE-311
Published
Modified
Record from
the NVD, synced

Affected

References

Check your stackJSON