cve.bz

CVE-2015-3953

9.8 criticalEPSS 2.0%

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Severity

CVSS 3.0

9.8 critical
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
2.0% chance of exploitation in the next 30 days; higher than 80% of CVEs (scores of 2026-10-31)
CWEs
CWE-798, CWE-259
Published
Modified
Record from
the NVD, synced

Affected

References

Check your stackJSON