cve.bz

CVE-2015-1012

7.5 highEPSS 0.8%

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Severity

CVSS 3.0

7.5 high
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS
0.8% chance of exploitation in the next 30 days; higher than 55% of CVEs (scores of 2026-10-31)
CWEs
CWE-200, CWE-312
Published
Modified
Record from
the NVD, synced

Affected

References

Check your stackJSON