cve.bz

CVE-2014-5434 baxter sigma spectrum infusion system firmware: Use of Hard-coded Credentials

9.8 criticalEPSS 1.6%

Critical (9.8) in baxter sigma spectrum infusion system firmware. EPSS 1.6% estimated probability of exploitation activity in the next 30 days.

On this page
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Severity

CVSS 3.0

9.8 critical
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
1.6% (unchanged from 1.6% a week ago; 1.6% 30 days ago) estimated probability of exploitation activity in the next 30 days; 76th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-798 Use of Hard-coded Credentials
CWE-259 Use of Hard-coded Password
OWASP Top 10
A07:2025 Authentication Failures
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2014-5434 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2014-5434?via=ask-ai-copy and https://cve.bz/cve/CVE-2014-5434?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2014-5434 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
baxter:sigma_spectrum_infusion_system_firmware6.05

ICS advisories (CISA)

AdvisoryTitlePublishedSectorsAffectedFixed
ICSA-15-181-01Baxter SIGMA Spectrum Infusion System Vulnerabilities2015-04-02Baxter SIGMA Spectrum Infusion System(model 35700BAX): 6.05none listed

From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.

Attack patterns (CAPEC)

  • CAPEC-70 Try Common or Default Usernames and Passwords Likelihood: Medium · Severity: High
  • CAPEC-191 Read Sensitive Constants Within an Executable Severity: Low

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON