cve.bz

CVE-2014-5434

9.8 criticalEPSS 1.6%

Critical (9.8) in baxter sigma spectrum infusion system firmware. EPSS 1.6% chance of exploitation in the next 30 days.

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Severity

CVSS 3.0

9.8 critical
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
1.6% chance of exploitation in the next 30 days; higher than 75% of CVEs (scores of 2026-10-31)
CWEs
CWE-798, CWE-259
Published
Modified
Record from
the NVD, synced

Affected

References

Check your stackJSON