cve.bz

CVE-2014-5431 baxter sigma spectrum infusion system firmware: Use of Hard-coded Credentials

6.8 mediumEPSS 0.4%

Medium (6.8) in baxter sigma spectrum infusion system firmware. EPSS 0.4% estimated probability of exploitation activity in the next 30 days.

On this page
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Severity

CVSS 3.0

6.8 medium
CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
0.4% (unchanged from 0.4% a week ago; 0.4% 30 days ago) estimated probability of exploitation activity in the next 30 days; 32nd percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
CWEs
CWE-798 Use of Hard-coded Credentials
CWE-259 Use of Hard-coded Password
OWASP Top 10
A07:2025 Authentication Failures
Published
Modified
NVD analysis
Analyzed, since modified
Record from
the NVD, synced

Ask an AI about this CVE

Show the promptExplain CVE-2014-5431 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2014-5431?via=ask-ai-copy and https://cve.bz/cve/CVE-2014-5431?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2014-5431 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.

Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.

Affected

Versions

From the NVD's CPE matches.

ProductAffected
baxter:sigma_spectrum_infusion_system_firmware6.05

Attack patterns (CAPEC)

  • CAPEC-70 Try Common or Default Usernames and Passwords Likelihood: Medium · Severity: High
  • CAPEC-191 Read Sensitive Constants Within an Executable Severity: Low

The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits.

References

Check your stackJSON