CVE-2014-5406 hospira lifecare pcainfusion firmware
hospira lifecare pcainfusion firmware: no CVSS score yet. EPSS 1.2% estimated probability of exploitation activity in the next 30 days.
On this page
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
- EPSS
- 1.2% (unchanged from 1.2% a week ago; 1.2% 30 days ago) estimated probability of exploitation activity in the next 30 days; 69th percentile. Scored 2026-11-22. A forecast for this CVE across the internet, not for your own systems.
- CWE
- CWE-345 Insufficient Verification of Data Authenticity
- OWASP Top 10
- A08:2025 Software or Data Integrity Failures
- Published
- Modified
- NVD analysis
- Analyzed, since modified
- Record from
- the NVD, synced
Ask an AI about this CVE
Show the prompt
Explain CVE-2014-5406 for an engineer deciding whether to act. Use https://cve.bz/api/public/cves/CVE-2014-5406?via=ask-ai-copy and https://cve.bz/cve/CVE-2014-5406?via=ask-ai-copy as the source of truth: what is affected and which versions fix it, whether it is in CISA KEV, public exploits, and its EPSS (an internet-wide 30-day estimate of exploitation activity, not a forecast for my systems). If they give no fixed versions, use the CVE record https://cveawg.mitre.org/api/cve/CVE-2014-5406 or the vendor advisory, not third-party mirrors. Tell me how to check whether I am affected and how to verify a fix. Cite the sources and say where they are uncertain.Opens a third-party assistant; its answer is its own, not cve.bz's. The prompt holds only this CVE's id and public links.
Affected
Versions
From the CNA's record in the CVE List.
| Product | Affected |
|---|---|
| Hospira LifeCare PCA Infusion System | through 5.0 |
ICS advisories (CISA)
| Advisory | Title | Published | Sectors | Affected | Fixed |
|---|---|---|---|---|---|
| ICSA-15-125-01B | Hospira LifeCare PCA Infusion System Vulnerabilities | 2015-02-05 | Hospira LifeCare PCA Infusion System: <=5.0 | none listed |
From CISA's ICS advisories (CSAF), provided under CISA's notification. Only the facts are shown; each advisory is on CISA's page. Credits.
Attack patterns (CAPEC)
The attack patterns MITRE's CAPEC relates to this record's weaknesses, as MITRE names them. Credits. More on the page of CWE-345 (12).
References
- http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2015/icsa-15-125-01.json
- https://www.cisa.gov/news-events/ics-advisories/icsa-15-125-01
- https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
- https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01