{"cve":"CVE-2025-15653","nvd":{"published":"2026-06-02T22:16:15.973","modified":"2026-07-22T19:10:00.120","status":"Awaiting Analysis","description":"Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.","cvss_v4":{"vector":"CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N","score":7.0},"cvss_v3":{"version":"3.1","vector":"CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H","score":6.8},"cwes":["CWE-668"],"references":["https://static.draeger.com/security/download/Product-Security-Advisory-25-349-Zeus.pdf","https://www.vulncheck.com/advisories/dr-ger-zeus-ie-anesthesia-workstation-usb-interface-privilege-escalation"],"products":[],"source":"nvd","synced_at":"2026-10-19T22:37:48.043517290+00:00"},"epss":{"cve":"CVE-2025-15653","score":0.00169,"percentile":0.06296,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}