{"cve":"CVE-2020-15482","nvd":{"published":"2020-08-26T16:15:12.510","modified":"2026-06-17T02:56:43.147","status":"Modified","description":"An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the admin account. This allows an attacker to gain root access to the device over the local network.","cvss_v4":null,"cvss_v3":{"version":"3.1","vector":"CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H","score":7.8},"cwes":["CWE-287","CWE-319"],"references":["https://payatu.com/advisory/unauthenticated-telnet-service-in-niscomed-patient-monitor","https://www.niscomed.com/multipara-monitor.html","https://payatu.com/advisory/unauthenticated-telnet-service-in-niscomed-patient-monitor","https://www.niscomed.com/multipara-monitor.html"],"products":["niscomed:m1000_multipara_patient_monitor_firmware"],"source":"nvd","synced_at":"2026-10-19T21:09:20.498348299+00:00"},"epss":{"cve":"CVE-2020-15482","score":0.002,"percentile":0.09954,"date":"2026-10-31"},"kev":null,"osv":[],"attribution":["This product uses the NVD API but is not endorsed or certified by the NVD.","EPSS scores: FIRST Exploit Prediction Scoring System, https://www.first.org/epss","Known exploited vulnerabilities: CISA Known Exploited Vulnerabilities Catalog, https://www.cisa.gov/known-exploited-vulnerabilities-catalog","Package advisories: OSV (https://osv.dev), from the GitHub Advisory Database, PyPA, Go, RustSec and other ecosystem databases, under CC-BY 4.0 or CC0","CVE records: the CVE Program's CVE List (https://www.cve.org), under its terms of use (https://www.cve.org/Legal/TermsOfUse). CVE is a registered trademark of The MITRE Corporation; this product is not affiliated with the CVE Program."]}